Diagnostic accuracy of a point-of-care antigen test for SARS-CoV-2 and influenza in a primary care population (RAPTOR-C19)
OBJECTIVES: Limited evidence exists for the diagnostic performance of point-of-care tests for SARS-CoV-2 and influenza in community healthcare. We carried out a prospective diagnostic accuracy study of the LumiraDx™ SARS-CoV-2 and influenza A or B assay in primary care. METHODS: Total of 913 adults and children with symptoms of current SARS-CoV-2 infection were recruited from 18 UK primary care practices during a period when Omicron was the predominant COVID variant of concern (June 2022 to December 2022). Trained health care staff performed the index test, with diagnostic accuracy parameters estimated for SARS-CoV-2 and influenza against real-time reverse-transcription PCR (rtRT-PCR). RESULTS: 151/887 participants were SARS-CoV-2 rtRT-PCR positive, 109 positive for Influenza A, 6 for Influenza B. Index test sensitivity for SARS-CoV-2 …